Menu
Index

System suitability criteria

 

5PL model

 
 
  Most of the suitability criteria options have an associated section that controls what action is taken if a suitability criterion fails.
 
For information only
 Yes / No
If  selected, this criterion will be marked "For Information Only" in the report. This option enables users to set criteria that they wish to track or investigate, but are not to be regarded as part of the regulatory report.
Continue assay on failure
 Yes / No
If selected, the report analysis will continue to the next steps even  if a criterion fails.
Print RP on failure
Yes / No
If selected, and the "continue" option is also selected, the report will print a value for the RP despite criteria failure.
 
System Suitability
Test group
Test
Criteria
Default values
Control Wells
 
 
 
 
 
Descriptive statistics
 
 
 
 
 
Mean of the group
Upper and lower limit
 
 
 
Mean of individual "lines"
 
 
 
 
Geometric mean of the group
 
 
 
 
Geometric mean of individual "lines"
 
 
 
 
 
 
 
Reference
Test group
Test
Criteria
Default values
 
Goodness of fit
 
 
 
 
Goodness of fit
p-value for F test if more than one replicate.
If there is only 1 replicate, there is no Goodness of Fit estimation. If GOF is selected in this situation the GOF test will be reported as  "unable to calculate", the estimate of the value will be "No result" and the pass / fail will be recorded as "FAIL"
 
Critical p value
Default p=0.01
 
 
 
 
 
 
Parameter values
 
 
 
 
Allowed ranges for parameters
A
Upper and lower limit
 
 
 
B
 
 
 
 
C
 
 
 
 
D
 
 
 
 
E
 
 
 
 
Location of midpoint (ED50)
 
 
 
 
Asymmetry
 
 
 
Equivalence tests for parameters
CI for A
Confidence level
Upper and lower limits 
 
 
 
CI for B
 
 
 
 
CI  for C
 
 
 
 
CI for D
 
 
 
 
CI for E
 
 
 
 
CI for Location of midpoint (ED50)
 
 
 
 
CI for Asymmetry
 
 
 
Significance of slope
p-value for B being significantly different from zero
Critical p-value.
 
 
Range
Limits for |D-A|
Upper and lower limits
 
 
 
CI for |D-A|
Confidence level
Upper and lower limits 
 
 
Variability of data
 
 
 
 
Variability
Variance estimate
Upper limit
 
 
 
RMSE
Upper limit
 
 
 
 
 
 
 
 
 
 
 
Quality Control
Test group
Test
Criteria
Default values
 
Goodness of fit
 
 
 
 
Goodness of fit
p-value for F test if more than one replicate.
If there is only 1 replicate, there is no Goodness of Fit estimation. If GOF is selected in this situation the GOF test will be reported as  "unable to calculate", the estimate of the value will be "No result" and the pass / fail will be recorded as "FAIL"
Critical p value
 
 
 
 
 
 
 
Parameter values
 
 
 
 
Significance of dose response
 
p-value for B being significantly different from zero
 
 
 
 
 
 
 
Parallelism tests
 
 
 
 
Ranges for combinations of parameters
Range for A difference
Upper and lower limits
 
 
 
Range for D difference
 
 
 
 
Range for Normalised A difference
 
 
 
 
Range for Normalised D difference
 
 
 
 
Range for B difference
 
 
 
 
Range for E difference
 
 
 
 
Range for Asymmetry  difference
 
 
 
 
Range Slope ratio
 
 
 
 Equivalence tests for combinations of parameters
CI for A difference
(Delta Method or Fieller's)
Confidence level for CI.
CI Upper and lower limits
 
 
 
CI for D difference
 
 
 
 
CI for Normalised A difference
 
 
 
 
CI for Normalised D difference
 
 
 
 
CI for B difference
 
 
 
 
CI for E difference
 
 
 
 
CI for Asymmetry difference
 
 
 
 
CI for Slope ratio
 
 
 
Significance tests
F test
Critical P value
 
 
 
Chi squared test
Critical P value
 
 
 
 
 
 
 
Relative potency tests
 
 
 
 
Allowed range
Relative Potency
Upper and lower limits
 
 
Equivalence test
Confidence interval for relative potency
Confidence level for CI.
CI Upper and lower limits
 
 
 
Precision
Ratio of the upper and lower ends of the CI for the RP, after back-transformation.
Confidence level for CI
upper limit for precision