Most of the suitability criteria options have an associated section that controls what action is taken if a suitability criterion fails.
For information only
Yes / No
If selected, this criterion will be marked "For Information Only" in the report. This option enables users to set criteria that they wish to track or investigate, but are not to be regarded as part of the regulatory report.
Continue assay on failure
Yes / No
If selected, the report analysis will continue to the next steps even if a criterion fails.
Print RP on failure
Yes / No
If selected, and the "continue" option is also selected, the report will print a value for the RP despite criteria failure.
System Suitability
Test group
Test
Criteria
Default values
Control Wells
Descriptive statistics
Mean of the group
Upper and lower limit
Mean of individual "lines"
Geometric mean of the group
Geometric mean of individual "lines"
Reference
Test group
Test
Criteria
Default values
Goodness of fit
Goodness of fit
p-value for F test if more than one replicate.
If there is only 1 replicate, there is no Goodness of Fit estimation. If GOF is selected in this situation the GOF test will be reported as "unable to calculate", the estimate of the value will be "No result" and the pass / fail will be recorded as "FAIL"
Critical p value
Default p=0.01
Parameter values
Allowed ranges for parameters
A
Upper and lower limit
B
C
D
E
Location of midpoint (ED50)
Asymmetry
Equivalence tests for parameters
CI for A
Confidence level
Upper and lower limits
CI for B
CI for C
CI for D
CI for E
CI for Location of midpoint (ED50)
CI for Asymmetry
Significance of slope
p-value for B being significantly different from zero
Critical p-value.
Range
Limits for |D-A|
Upper and lower limits
CI for |D-A|
Confidence level
Upper and lower limits
Variability of data
Variability
Variance estimate
Upper limit
RMSE
Upper limit
Quality Control
Test group
Test
Criteria
Default values
Goodness of fit
Goodness of fit
p-value for F test if more than one replicate.
If there is only 1 replicate, there is no Goodness of Fit estimation. If GOF is selected in this situation the GOF test will be reported as "unable to calculate", the estimate of the value will be "No result" and the pass / fail will be recorded as "FAIL"
Critical p value
Parameter values
Significance of dose response
p-value for B being significantly different from zero
Parallelism tests
Ranges for combinations of parameters
Range for A difference
Upper and lower limits
Range for D difference
Range for Normalised A difference
Range for Normalised D difference
Range for B difference
Range for E difference
Range for Asymmetry difference
Range Slope ratio
Equivalence tests for combinations of parameters
CI for A difference
(Delta Method or Fieller's)
Confidence level for CI.
CI Upper and lower limits
CI for D difference
CI for Normalised A difference
CI for Normalised D difference
CI for B difference
CI for E difference
CI for Asymmetry difference
CI for Slope ratio
Significance tests
F test
Critical P value
Chi squared test
Critical P value
Relative potency tests
Allowed range
Relative Potency
Upper and lower limits
Equivalence test
Confidence interval for relative potency
Confidence level for CI.
CI Upper and lower limits
Precision
Ratio of the upper and lower ends of the CI for the RP, after back-transformation.