1) Fit a parallel model to reference standard and test sample pair:
  and 
 
2) Calculate Log (RP) = (ATest -  ARef   ) / B
3) Back-transform using the same log base as was used for the dose:
 

Confidence interval for relative potency

The variance, s2, of the estimate of Log (RP) is calculated from the variances and covariances of the model parameter estimates of ,    and  using the Delta method. 
The 95% confidence interval for Log (RP) is then calculated as:
 
·     Log (RP)  ± z 0.025 √ s2 
Where:
 
 z 0.025 is the 97.5% percentile of the normal distribution.
·     The 95% confidence interval for RP is obtained by back-transforming the interval for Log RP using the same log base as was used for the dose.